Lot 59Olympus CF-P20L Colonoscope in Case - Engineer's Report : Optical System - Approximately 120 Broken Fibers and Bad Fluid Stain Present, Angulation - Not Reaching Specification, Insertion Tube - Split, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leaking *2910407*